Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5500 DATE: 22-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 18 NUMBER OF LINES: 181 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 RSLT =CHANNEL [MEM] 1.004 HEAD CHANNEL [MEM]: {AMP MEASUREMENT} 1.005 DISP Connect the 5500 NORMAL OUTPUT to the CH[MEM] INPUT. 1.006 DISP Select the DIRECT LOG mode. 1.007 DISP Set the following parameters in the PLOT menu... 1.007 DISP 1.007 DISP INTERVAL to 1s 1.007 DISP CH. SELECT to Appropriate Channel Bank 1.008 DISP Using the INITIALIZE 1 (ETC,ETC,INI1,F2) key, set the 1.008 DISP position to 100%. Press SLCT key to enter the value. 1.008 DISP 1.008 DISP Select the DISPLAY mode and set the position to -2% 1.008 DISP by pressing the POSI key and then pressing the 1.008 DISP POSITION keys for CH[MEM]. 1.009 DISP Select the DIRECT LOG mode. 1.009 DISP 1.009 DISP Set the FILTER to 5Hz and MEAS. to DC. 1.010 RNG 500mV 1.011 DISP Press the START/STOP key. 1.012 5500 500mV S 2W 1.013 MATH S[1] = "Measurement is 500mV +/-1.2mV at 100% position." 1.014 EVAL -s S[1] to Is the measurement 500mV +/-1.2mV at 100% position? 2.001 DISP Using the INITIALIZE 1 (ETC,ETC,INI1,F2) key, set the 2.001 DISP position to 300%. Press SLCT key to enter the value. 2.001 DISP 2.001 DISP Select the DISPLAY mode and set the position to -2% 2.001 DISP by pressing the POSI key and then pressing the 2.001 DISP POSITION keys for CH[MEM]. 2.002 DISP Select the DIRECT LOG mode. 2.002 DISP 2.002 DISP Set the FILTER to 5Hz and MEAS. to DC. 2.003 RNG 200mV 2.004 5500 200mV S 2W 2.005 MATH S[1] = "Measurement is 200mV +/-1.5mV at 300% position." 2.006 EVAL -s S[1] to Is the measurement 500mV +/-1.5mV at 300% position? 3.001 DISP Press the START/STOP key. 3.002 ASK- N 3.003 HEAD CHANNEL [MEM]: {RANGE SWITCHOVER} 3.004 DISP Select the DIRECT Y-T mode. 3.005 DISP Set the following parameters in the PLOT menu... 3.005 DISP 3.005 DISP CHART SPEED: 2mm/s 3.005 DISP FORMAT: 200mm x 1 3.006 DISP Using the INITIALIZE 1 (ETC,ETC,INI1,F2) key to set the 3.006 DISP position to 100%. 3.006 DISP 3.006 DISP Select the DISPLAY mode and set the position to 0% 3.006 DISP by pressing the POSI key and then pressing the 3.006 DISP POSITION keys for CH[MEM]. 3.007 DISP Select the DIRECT Y-T mode. 3.007 DISP 3.007 DISP Set the FILTER to 5Hz and MEAS. to DC. 3.008 RNG 500mV 3.009 DISP During the following tests, the difference between 3.009 DISP each range is compared to the reference range (500mV) 3.009 DISP and must be within +/-0.25%. 3.010 DISP Press the START/STOP key. 3.011 5500 500mV S 2W 3.012 DISP Note the full scale measurement (REFERENCE). 3.013 5500 * S 3.014 MESS Adjust stimulus for a full scale measurement that is the 3.014 MESS same as the full scale measurement on the 500mV Range. 3.015 5500 50 50mV 0.25% 2W 4.001 5500 100 100mV 0.25% 2W 5.001 5500 200 200mV 0.25% 2W 6.001 5500 1 1V 0.25% 2W 7.001 5500 2 2V 0.25% 2W 8.001 5500 5 5V 0.25% 2W 9.001 DISP Press the START/STOP key. 9.002 ASK+ N 9.003 HEAD CHANNEL [MEM]: {MEMORY TRIGGER OPERATION CHECK} 9.004 DISP Select the DISPLAY mode and set the position to 50% by 9.004 DISP pressing the POSI key and then pressing the 9.005 DISP POSITION keys for CH[MEM]. 9.006 DISP Select the MEMORY Y-T mode. 9.007 DISP Set the following parameters in the PLOT menu... 9.007 DISP 9.007 DISP FORMAT to 20mm x 8 9.007 DISP GRID to 10mm Fine 9.007 DISP PAPER FORMAT to Normal 9.007 DISP SCALING FACTOR to 1 9.007 DISP LIST to No 9.007 DISP SCALE to Yes 9.008 DISP Set the following parameters in MEMORY menu... 9.008 DISP 9.008 DISP SAMP. INTV. to 10us 9.008 DISP MEMORY BLOCK to 16CHx1 9.008 DISP MEMORY BANK to 8k-1 9.008 DISP MEMORY SIZE to 12.5% 9.008 DISP PR. AREA to 0 to 100% 9.009 DISP Set the following parameters in TRIGGER menu... 9.009 DISP 9.009 DISP MODE: A 9.009 DISP CH[MEM] to 50% POSITIVE SLOPE 9.009 DISP TRIG. DELAY to +000% 9.009 DISP TRIG. ACTION to SINGLE 9.010 DISP Set the RANGE to 5 V. 9.010 DISP Set the FILTER to OFF and MEAS. to AC. 9.011 5500 1V 1kH SI S 2W 9.012 MESS Press the START/STOP key (stores the data), then 9.012 MESS press the PRINT key. 9.013 MATH S[1] = "Data stores and prints (POS SLOPE)." 9.014 EVAL -s S[1] : Does data store and print (POS slope)? 10.001 DISP In the TRIGGER menu, set to 50% NEGATIVE SLOPE. 10.001 DISP 10.002 5500 1V 1kH SI S 2W 10.003 MESS Press the START/STOP key (stores the data), then 10.003 MESS press the PRINT key. 10.004 MATH S[1] = "Data stores and prints (NEG SLOPE)." 10.005 EVAL -s S[1] : Does data store and print (NEG slope)? 11.001 MESS 11.002 HEAD CHANNEL [MEM]: {NOISE CHECK} 11.003 DISP Disconnect the 5500A from the UUT and short the 11.003 DISP CH[MEM] (+) and (-) input terminals. 11.004 DISP Select the DIRECT Y-T mode. 11.005 DISP Use the INITIALIZE (ETC,ETC,INI1,F2) key 11.005 DISP to select the 100% position. 11.005 DISP 11.005 DISP Select the DISPLAY mode and set the position to 50% 11.005 DISP by pressing the POSI key and then pressing the 11.005 DISP POSITION keys for CH[MEM]. 11.006 DISP Select the DIRECT Y-T mode. 11.007 DISP Set the following parameters in the PLOT menu... 11.007 DISP 11.007 DISP FORMAT to 200mm x 1 11.008 DISP Set the following parameters in TRIGGER menu... 11.008 DISP 11.008 DISP MODE to MANUAL 11.009 DISP Set the FILTER to OFF and MEAS. to DC. 11.010 DISP Press the START/STOP key. 11.011 RNG 0.05V 11.012 EVAL -s "Noise <= 1mm on 50mV Range." : Is noise less <= 1mm? 12.001 RNG 0.1V 12.002 EVAL -s "Noise <= 1mm on 100mV Range." : Is noise less <= 1mm? 13.001 RNG 0.2V 13.002 EVAL -s "Noise <= 1mm on 200mV Range." : Is noise less <= 1mm? 14.001 RNG 0.5V 14.002 EVAL -s "Noise <= 1mm on 500mV Range." : Is noise less <= 1mm? 15.001 RNG 1V 15.002 EVAL -s "Noise <= 1mm on 1V Range." : Is noise less <= 1mm? 16.001 RNG 2V 16.002 EVAL -s "Noise <= 1mm on 2V Range." : Is noise less <= 1mm? 17.001 RNG 5V 17.002 EVAL -s "Noise <= 1mm on 5V Range." : Is noise less <= 1mm? 18.001 DISP Press the START/STOP key. 18.002 END